In order to obtain reliable measurements calibrate on the same type of
base as the parts that you will actually be measuring.
Highest accuracy is obtained when you are closest to the calibration
point. This means that whenever possible, you should use shims that
are as close as possible to the thickness that you expect to measure.
Use the probe gently over the test piece that you will be measuring.
There is no need to press hard.
Do not use lubricants on the probe or any of its components and clean
using a dry cloth.
Avoid measuring edges to minimize errors. It is always best to measure
away from test piece edges.
When measuring very thin substrates or irregular surfaces, avoid using
“1 point” and “Only Base” calibration modes.
Shims can be overlapped (placed over one another) in order to obtain
thicker values. But labels should not overlap and should remain visible
at all times.
Correct
28
Incorrect