7.2.2 Frequency controls
The area below the trace screen contains all the controls for setting the frequency range to be scanned
and/or single frequency controls.
7.2.2.1 Steps/scan
Controls the frequency resolution of the analyser. This sets the number of frequency points measured
per scan. Note that whilst in Linear sweep mode, the spacing is linear, in Log sweep mode the spacing
is set to give constant screen resolution and a true log scan. The maximum number of steps is 500.
Selecting any higher number would have no benefit as there are only 500 pixels across the trace screen.
7.2.2.2 Frequency range
These set the frequency range to be scanned. Low range covers 10KHz to 30MHz.
High range covers 30MHz to 1100MHz. The selected range is shown just to the right of the buttons.
Changing the frequency range causes the analyser to enter a self calibration mode which takes 30
seconds.
The I.F. bandwidth is automatically changed when the range is changed. For Low range, I.F. bandwidth
is 9KHz. For High range, I.F. bandwidth is 120KHz.
7.2.2.3 Frequency start/finish
These set the start (low frequency end) and stop (high frequency end) frequencies of
the scan. They cause a frequency entry window to appear enabling the chosen frequency to be keyed in
MHz. If the Start button is clicked, the software will automatically show the Stop entry window after
the Start frequency has been entered.
On the low frequency range, the entry is XX.XX MHz. On the high frequency range, only integer
numbers are allowed.
Software logic will ensure that the 30MHz boundary is not crossed.
Note that when in Log sweep mode the frequency start/stop parameters cannot be changed.
A full list of techniques for setting the scan range is as follows:
1. The cursor can be used to zoom into a narrower frequency range than currently displayed. This
technique works with Lin frequency axis only, not Log. The system works by using the cursor to
’pull± each edge to the desired position. To do this, position the cursor on the trace screen within 1
division of the LH edge, hold the left mouse button down and drag the LH edge to the desired
frequency then release the button. Repeat for the right hand edge of the trace screen.
2. The keyboard can be used to type in the desired start and stop frequencies by clicking on either the
Start or Stop buttons.
3. The frequency axis can be ’shifted± left or right by half the screen width using the
buttons immediately below the trace screen.
4. The frequency axis can be ’zoomed out± by clicking on the
the frequency coverage will double.
or
button. Each time this is clicked
5. The frequency axis can be reset to the full frequency span (30 - 1100MHz or 10KHz - 30MHz) by
clicking on the Low or High buttons.
6. If in Log frequency axis,
or
will switch between alternative ranges.
If in High range, the alternatives are 30 - 300MHz span and 30 - 1000MHz span.
For low range the corresponding alternatives are 10KHz - 30MHz and 100KHz - 30MHz.
29