Construction manual RoeTest V8 - (c) Helmut Weigl
Page 56
The output voltage is calibrated for both the low and high ends of the output range. But
because the D/A converters are less accurate towards the far ends of the range (digital 0
and 255) we don’t calibrate at the extreme ends, but use the D/A values of 20 and 230 to
calibrate.
Set the anode/plate voltage to low range (0-51V) and select a current DAC value of 20
either by moving the slider or by entering the value in the field where the DAC value is
displayed.
As an alternative the 20 or 230 DAC values for calibration can be selected by using the
buttons circled in red in the picture below:
Adjust the trim pot marked "unten, kleiner Bereich" on the anode PCB until your multimeter
indicates the same voltage as the value shown in red on the screen.
Select the anode/plate voltage high range and calibrate using the same procedure.
Then enter a DAC value of 230 and adjust the trim pot marked "oben, kleiner Bereich"until
the measured voltage equals the voltage now displayed in red on your computer screen.
Since the trim pot adjustments are somewhat interactive, repeat this calibration sequence a
few times. Select the anode/plate voltage high range and calibrate using the same procedure.
Calibrate the G1, G2, G3 and heater boards in a similar manner.