Download User manual

Transcript
11.6 ANALYSIS AND INTERPRETATION OF QUALITATIVE
RESULTS
At the end of the reaction, view the graph in logarithmic scale.
Analyze AML1-ETO and ABL graphs and amplification results separately and
follow the interpretation pattern as indicated.
Before analyzing the samples results check the expected results of the
positive and negative controls.
ABL
positive Control
ABL
negative Control
inv16
positive Control
inv16
negative Control
RESULT
INTERPRETATION
Amplification signal
Correct ABL amplification
No amplification signal
Amplification
problems,
repeat the analysis
No amplification signal
No contamination
Amplification signal
Contamination,
analysis
RESULT
INTERPRETATION
Amplification signal
Correct
amplification
No amplification signal
Amplification
problems,
repeat the analysis
No amplification signal
No contamination
Amplification signal
Contamination,
analysis
19
repeat
the
AML1-ETO
repeat
RQ-S61-48_EN.doc
the