Model 5014i to 5030i Upgrade Procedures Radius Tube Adapter Installation CONNECT TO SHARP 1 2-PIN IS NOT USED Figure 10–4. Connecting Connection Bracket Wiring to Measurement Interface Board Radius Tube Adapter Installation Use the following procedure for radius tube adapter installation. 1. Loosen the Radius Tube Mounting bracket screws. 2. Slide back bracket downwards. 3. Slowly remove the existing Radius Face Seal by pulling straight up with a twisting action. 4. Slowly insert the Radius Tube Adapter onto the Radius Tube and push down with a slight twist until it is seated. 5. Tighten the radius tube mounting bracket screws as shown in Figure 10–5. Thermo Fisher Scientific Model 5014i Beta Instruction Manual 10-5