This User Manual contains information on design, principle of operation, characteristics of the
product and instructions essential for correct and safe use of this product (intended use, maintenance,
servicing, storage and transportation), as well as information regarding the utilization of the product.
1 DESCRIPTION AND OPERATION OF THE PRODUCT
1.1 Product functionality
Wide range gamma area monitor DBG-S11D are designed for continuous measurement of
ambient dose equivalent rate of gamma radiation (hereinafter – ADER).
Dosimeters are used for radiation monitoring at industrial facilities, nuclear plants, and
enterprises for radioactive waste conditioning and management and nearby areas.
Dosimeters can be used as standalone units or as part of a systems, complexes and installations
for radiation monitoring.
Dosimeters are available in three modifications, which differ by measurement range and
dimensions.
1.2 Technical characteristics
1.2.1 Energy range of measured gamma radiation …………………….……… 0.05 to 3.0 MeV.
1.2.2 Measurement range of the ADER of gamma radiation:
- basic modification ……………………………………..…..……… 0.1 μSv·h-1 to 10 mSv·h-1;
- modification 01 ……………………………………………………… 0.1 μSv·h-1 to 10 Sv·h-1 ;
- modification 02 ………………………………………………..…… 0.1 μSv·h-1 to 100 Sv·h-1.
1.2.3 Limits of the permissible basic relative measurement error of ADER of gamma radiation:
- in the range 0.1 μSv·h-1 to 1 μSv·h-1 …………………...……………………… ± (15+1/H) %;
- where H is a dimensionless quantity numerically equal to the measured value of the ADER
of gamma radiation expressed in μSv·h-1
- in the range 1 μSv·h-1 to 10 Sv·h-1 …………….….…………………………………… ±15 %;
- in the range 10 μSv·h-1 to 100 Sv·h-1 ……………….…….…………………………… ±20 %.
1.2.4 Energy dependence of sensitivity of dosimeters relative to 137Cs (662 keV) …...… ±25 %.
1.2.5 Dosimeter warm-up time …………………………………………………………… 1 min.
1.2.6 Continuous operation without limitation of number of turning on/off …..……… 24 hours.
1.2.7 Instability during 24 hours of continuous operation relative to the mean value
of readings for this period …………………………………….………………………………… ±10 %.
1.2.8 Dosimeters are capable to transfer current value of measured quantity, estimation of its
statistical uncertainty and self-diagnostics data to the external information network in accordance with
the DiBUS protocol using the interfaces RS-485, USB.
Description of data exchange registers of the DiBUS protocol is given in Appendix E.
1.2.9 Power supply voltage ………………………………………..………….………… 12 63 V.
1.2.10 Current consumption at voltage 12 V ..………..……………….………..……….. 50 mA.
1.2.11 Operating conditions:
- operating temperature …………………………………………………… minus 60 to +80 °C;
- relative humidity …………..………………..…………………………..…… 98 % at +35 C;
- atmospheric pressure ………………………………………...………….… 84.0 to 106.7 kPa;
- content of the corrosive agents in the ambient air corresponds to the values in Table 1.1.
FVKM.412113.042RE
3