Download User Manual

Transcript
5. Operation
These instructions cover day-to-day operation of the system. Other operations,
such as maintenance and repair, should only be carried out by suitably qualified
personnel.
System
Optimisation
The Autoexcel system automatically compensates for
patient weight distribution and position optimising the
pressure relieving performance. To ensure that the
pressure relieving properties are not impaired, the
cover must not be pulled tight and any covering sheets
are not fitted too tightly.
Silencing
Audible Alarms
Audible alarms can be silenced using the MUTE
pushbutton. To silence an alarm simply push the MUTE
button once.
CPR Control
IMPORTANT
In the event of cardiac arrest, disconnect the tubeset connector
at the foot end of the overlay.
A swing label indicates the position of the connector in
the event of bed covers obscuring the tubeset location.
+
The overlay will deflate sufficiently for CPR procedures to
commence within 10 seconds of disconnection.
12