Transcript
DOCUMENT 301925, REVISION A Figure 9 shows a sample of the most common setting, appropriate for many strong motion applications when the EpiSensor is used with an Altus recorder. Figure 9: Jumper settings for 2g, 2.5V, single -ended configuration Note that while the above settings are independent, each of the options must be configured correctly for the EpiSensor to function properly. Power Supply Options The EpiSensor is offered in two power supply configurations: § § Dual-supply (standard) Single-supply (optional) The jumpers for these options are set at the factory and should not be changed. Caution: Damage to instrument. When the EpiSensor is configured for single-supply operation do not connect any power source to the brown wire or Pin H (on the connector). Calibration Coil The EpiSensor is equipped with a calibration coil. This coil is isolated from other EpiSensor circuitry and accurately emulates the effect of acceleration on the system. This coil can be used to verify both the static acceleration sensitivity of the EpiSensor and the dynamic response of the system. When used with Altus recorders, the calibration signals are automatically applied EPI SENSOR ES-U U SER G UIDE 23