Download 3DSpaceDx

Transcript
The Device Status area will show how many sensors are detected (4), and the XYZ coordinates of the stylus. The rest of the display contains 3 columns: Markers, Electrodes, and Head
Shape. In this tutorial, we will be using a model head with the measured electrode positions
marked on the head. This will allow us to digitize the PAN coordinate markers, the electrodes,
and the head shape in that order. In reality, you will likely perform the measurements in a
different order. For example, you might digitize the PAN coordinates, then digitize the head
shape. After that you would typically then put on an electrode cap, replace the sensors in their
precise, original locations, and digitize the electrode positions (described in more detail below).
This sequence allows you to digitize the head shape without having to go around the electrodes.
To make the PAN measurements as accurate as possible, you should mark the Nasion, Left
and Right measurement points with a wax pencil so you can identify the precise locations when
you replace the sensors. If you plan on acquiring data from the same subject at separate points
in time (across days), it would be beneficial to take pictures showing the precise marker
measurements. It is very important to obtain as identical as possible marker measurements.
It is necessary to digitize the PAN coordinates first. These are the left and right preauricular
points and the nasion. These are the 3 required points to form the reference plane. Place the
stylus tip at the left preauricular point (detailed below) and click the button on the stylus one
time. You should see a "1" by the Left button when the point has been digitized. Repeat the
same operations for the Nasion and Right preauricular points. If you make a mistake, you can
click the Clear button to start over.
You can obtain repeated measurements of the 3 points for more precise calculations (it is a
good idea to digitize the points 3 time each). Note that the program will cycle from Left to
Nasion to Right repeatedly. You also have the option to digitize the Inion and ZVertex, if so
desired. These points may be used by other source localization software, such as Curry. They
are not required for 3DSpaceDx.
As you digitize the PAN coordinates, you will see the points appear on the screen (L, R and
N). You can enlarge the display by clicking the Enlarge icon
appear similar to the following.
. Your display now may
3DSpaceDX - 21
Copyright © 2001 Neurosoft, Inc.
Revision B, Document # 2201