11 THEORY OF OPERATION and MATHEMATICAL ANALYSIS When the high-energy electrons emanating from the radioactive decay of 14C (carbon-14) interact with nearby matter they loose their energy and, in some cases, are absorbed by the matter. These high-energy electrons emitted through radioactive decay are known as beta rays and the process is known as beta-ray attenuation. When matter is placed between the radioactive 14C source and a device designed to detect beta rays, the beta rays are absorbed and/or their energy diminished. This results in a reduction in the number of beta particles detected. The magnitude of the reduction in detected beta particles is a function of the mass of the absorbing matter between the 14C beta source and the detector. The number of beta particles passing through absorbing matter, such as dust deposited on a filter tape, decrease nearly exponentially with the mass through which they much pass. Equation 1 shows this relationship. Equation 1 I = I 0 e−µx In Equation 1, I is the measured beta ray intensity (counts per unit time), of the attenuated beta ray (dust laden filter tape), I0 is the measured beta ray intensity of the un-attenuated beta ray (clean filter tape), µ is the absorption cross section of the material absorbing the beta rays (cm2/g), and x is the mass density of the absorbing matter (g/cm2). Equation 1 very closely resembles the Lambert-Beers Law, which is used in spectrometric analysis. Just as the Lambert-Beers Law is an idealization of what is actually observed, Equation 1 is also an idealized simplification of the true processes occurring meant to simplify the corresponding mathematics. However, experimental measurement shows that in properly designed monitors, such as the BAM-1020, the use of this equation introduces no substantial error. Equation 1 may be rearranged to solve for x, the mass density of the absorbing matter. This is shown in Equation 2. Equation 2 − 1 I 1 I0 ln = ln =x µ I 0 µ I In practice, the absorption cross section is experimentally determined during the calibration process. Once I and I0 are experimentally measured, it is a simple matter to calculate x, the predicted mass density. In practice, ambient air is sampled at a constant flow rate (Q) for a specified time ∆t. This sampled air is passed through a filter of surface area A. Once x, the mass density of collected particles, has been determined, it is possible to calculate the ambient concentration of particulate matter (µg/m3) with Equation 3. BAM-1020-9800 REV K Page 105