Section 3: System Operation
2 Use the contact button on the scan toolbar. This is on the right hand
side of the toolbar. The controller is set for contact when the button is
depressed.
3.5.3 Setting up the Controller for Non-Contact Tuning
Before continuing, you should:
1 Make sure that the tip is away from the sample so that it is not
affected by the surface. A tiny visible gap is sufficient.
2 Set the reference amplitude to zero. Switch the SPM MONITOR on the
NSOM TOPAZ controller to the REFERENCE FORCE AMPLITUDE /
CURRENT position and adjust the REFERENCE FORCE
AMPLITUDE / CURRENT setting to give a reading of zero.
3 Set the oscillation amplitude to about one volt. Switch the SPM
MONITOR to the OSCILLATION AMPLITUDE position and adjust
with the BIAS VOLTAGE / OSCILLATION AMPLITUDE knob.
3.5.4 Setting the Non-Contact Tuning Parameters in Quartz
Now the hardware is configured click on the Calibrate menu, then
on the Non-Contact Tuning sub-menu and finally on the Scan command.
Alternatively, use the button on the toolbar. A non-contact tuning window
will appear and you will be presented with a dialog box for setting the
scan parameters. This is shown below.
Figure 3.2: The Non-Contact Tuning Dialog
1 Frequency Range: The NSOM TOPAZ controller has two frequency
ranges. In the low range the frequency can be varied between 0 and
100 kHz and in the high frequency range it can be scanned between 80
and 480 kHz. First use the High Frequency Range check box to select
21