Download HI 9829 - Hanna Instruments
Transcript
• Unscrew the calibration beaker and empty the solution. • Shake any remaining liquid off the probe and beaker. No droplets should remain on the D.O. sensor membrane. Note Do not attempt to dry wipe the D.O. sensor as damage to the mem brane may occur. • Screw the empty calibration beaker on the probe body. The beaker should not be dry. • Press <Accept> to close the displayed message. • When the measurement is stable, “Ready” appears. Press <Confirm> to store the calibration data and the “Storing” message will appear. • Press <OK> to return to “Calibration” menu. Note To quit the quick calibration procedure, press ESC at any time. After every calibration the quick calibration window will show a check mark in the box next to the calibrated parameter. 37