Download HI 3512 - Hanna Instruments
Transcript
PARx sign and decimal point (6 chars) • Salinity calibration data (if available), which contains: • the number of calibrated standards (1 char always 1) • the salinity factor, with sign and decimal point (7 chars) • calibration time, yymmddhhmmss (12 chars) Requests the setup parameters setting. • x=1 setup parameters for channel 1 • x=2 setup parameters for channel 2 The answer string contains: • Lot log interval index (2 ASII chars - it represents the index from the setup list for the log interval) • Backlight Value (1 ASCII char) • Contrast Value (2 ASCII char) • Instrument ID (4 chars) Channel 1 data • Calibration Alarm time out for pH (2 chars) • Calibration Alarm timeout for ISE (2 chars) • SETUP information (2 chars): 8 bit hexadecimal encoding. • 0x01 - beep ON (else OFF) • 0x02 - degrees Celsius (else degrees Fahrenheit) • 0x04 - Offset calibration (else Point calibration) • The number of pH custom buffers (1 char) • The custom buffer values, with sign and decimal point, for each defined custom buffer (7 chars) • The ID of the ISE electrode (2 chars) (standard or custom) • The molar weight of the selected ION, with sign and decimal point (9 ASCII characters) • The electrode slope (6 ASCII chars) • The ion charge (2 chars) • The ISE unit (2 chars) • The short name of the selected language (3 chars) Channel 2 data • Calibration alarm timeout for the EC (2 chars) • Calibration alarm timeout for the Salinity (2 chars) • Setup information (2 chars): 8 bit hexadecimal 66