Download Microstructure Profiler MSS90

Transcript
For data evaluation, the program DatPro can be used. DatPro is
specially designed for customers of the MSS Profiler. It has the character
of a toolbox. It consists of a management program and many modules to
carry out various steps of data evaluation. The management program
handles all modules and manages the dialogue with the user of the
program. The user of DatPro can select the various steps and the
sequence in the data evaluation process dependent on his needs and
aims of his investigation. Single modules can be combined to macros.
The macros can be applied to one data file or to a list of data files.
Graphical utilities enable a comfortable quick-look of the measured and
processed data.
The MSS microstructure measuring system has some outstanding
properties, which results to a superior performance related to other
microstructure profilers. The MSS90 profiler has a high sample rate and
high analogue band width. The system allows a number of sensors to be
connected directly to the bottom of the profiler without the need to use
underwater connections. The following table lists the most important
features in a short review:
- Sample rate:
- Number of parameters
- number of analogue channels:
- Bandwidth of microstructure channel:
- response time of MS channels
- Resolution (all channels):
1024 data sets/sec
16 / dataset
15
150 Hz (-3dB)
< 12 ms
16 bit
Data transmission is 614,4 kbaud (tested for cable length up to 1000 m).
1.2 Sensor equipment
The MSS90 profiler may be equipped with up to 9 different sensors on
the bottom cap, which can be subdivided in two different groups:
- Standard CTD sensors
- Microstructure sensors
The standard CTD sensors have a relatively slow response but high
accuracy. The following sensors are available:
- Pressure
- Temperature (PT 100)
- Conductivity