Download Description - Agilent Technologies

Transcript
CALL:REGister
CALL:REGister:TIMer[:SRPeriod]
Function
Sets the time between time based registrations (when timer based registration is on) and turns
the timer based registration state to on.
The value is not the actual time. The registration period is used in the following equation to
determine the time between registrations:
time between registrations= [integer portion of 2(reg_period/4)] x 0.08 seconds
Setting
Range: 29 to 85
Resolution: integer
Query
Range: 29 to 85
*RST Setting
29 (12 seconds)
Programming Example
OUTPUT 714;"CALL:REGISTER:TIMER:SRPERIOD 35"
!Turns the timer’s state on, and sets the registration period to 35.
CALL:REGister:TIMer:STATe
Function
Sets the state of timer based registrations.
Setting
Range: 1|ON|0|OFF
Query
Range: 1|0
*RST Setting
ON
Programming Example
OUTPUT 714;"CALL:REGISTER:TIMER:STATE OFF"
CALL:REGister:TIMer:RPERiod
Function
Sets the time between time based registrations (when timer based registration is on). The value
is not the actual time. The registration period is used in the following equation to determine the
time between registrations.
time between registrations = [integer portion of 2(reg_period/4)] x 0.08 seconds
Setting
Range: 29 to 85
Resolution: integer
Query
Range: 29 to 85
*RST Setting
29 (12 seconds)
Programming Example
OUTPUT 714;"CALL:REGISTER:TIMER:RPERIOD 35"
261
S:\Hp8960\E1962B CDMA 2000\3.1_reference_guide\Chapters\cdma2000_hpib_call_register.fm