Download GasAlertMicroClip Technical Reference Guide
Transcript
GasAlertMicroClip Troubleshooting Table 15. Troubleshooting Problem Possible Cause Solution Detector Operation Detector does not display expected gas readings after activation self-test. Detector does not respond to pushbutton. Detector does not accurately measure gas. Detector does not enter into alarm. Sensor not stabilized Used sensor: wait 60 seconds New sensor: wait 5 minutes Sensor(s) requires calibration Calibrate the sensor(s). Refer to Calibration. Target gas is present Detector is operating properly. Use caution in suspect areas. Battery is depleted Charge battery. Refer to Charging the Battery. Detector is performing operations that do not require user input Pushbutton operation restores automatically when the operation ends. Sensor(s) requires calibration Calibrate the sensors. Refer to Calibration. Detector is colder/hotter than ambient gas Allow the detector to attain ambient temperature before use. Sensor filter is blocked Replace the sensor filter. Refer to Replacing a Sensor or Sensor Filter. Alarm setpoint(s) defined incorrectly Reset alarm setpoints. Refer to Factory Gas Alarm Setpoints and Sensor Configuration. Alarm setpoint(s) set to zero Reset alarm setpoints. Refer to Factory Gas Alarm Setpoints and Sensor Configuration. Detector is in calibration mode Complete the calibration procedure. 53