Download User`s Manual - Stanford Research Systems

Transcript
PC Applications
132
• Axis linewidth: the width of the box surrounding the plot, in pixels.
• Grid linewidth: the width of the plot gridlines, in pixels.
• Plot linewidth: the width of the plot traces, in pixels. Values other than 1 may
significantly increase the amount of tie that it takes to draw the graph.
Show statistics
Shows information such as the average, minimum, and maximum values for all data within the
graph’s X range. Only information for the buffer plotted in black is shown.
Linear regression
The linear regression feature can be used to determine how much one temperature sensor is
miscalibrated compared to another. You are asked to choose an X and a Y buffer (the log files for
two different temperature sensors). The software then determines the offset and gain of the X
buffer relative to the Y buffer. Check the "Apply equation to X buffer" box to multiply the X
buffer by the gain factor and then add the offset.
Command line
Opens a File Grapher command line window. The commands described in the table below can
be typed into the command line. Sequences of commands can be stored as macros and then
recalled either from the command line or the Special menu.
Align X axes
Sets the X axis range of all graphs to be equal to the X axis range of the selected graph.
Add graph
Adds another graph to the display. When more than one graph is displayed, you can select a
graph by clicking on it. Most operations only apply to the selected graph.
Overall plot size
Changes the size of the entire plot window and all the graphs in the window.
Set as baseline
When this option is selected, the channel that is currently displayed becomes the baseline and is
subtracted from all displayed data. Selecting this option does not modify the data in any way, just
the way the data is displayed.
Clear baseline
Disables the baseline feature. This option is grayed out if no baseline is currently set.
Subtract average
When selected, each file’s data is displayed with its average subtracted. Selecting this option
does not modify the data in any way, just the way the data is displayed.
Process menu
The process menu lets you modify data. The operations are applied to an internal copy of the
data (i.e., a buffer) and do not affect log files on disk. When you select an item from the process
menu, a dialog may appear asking which of the currently-plotted buffers you’d like to apply the
operation to.
PTC10 Programmable Temperature Controller