Download - Frank`s Hospital Workshop

Transcript
Nemko ref. no.:NK-2E-M-055
42
EXCESSIVE TEMPERATURES
42.1 - 42.2
Determination of the temperature with
thermocouples:
P
No. 42
Equipment:
Quantity
Rated
Used in tests
100-240V AC,
50/60Hz
264 V AC / 50Hz
Supply power, VA
—
—
Output power, VA
—
—
Supply voltage, V
Measuring point
Remarks
Blocked opening
Maximum normal load :
Refer to sub-clause 7(power input)
Measured
temp.
∆t (°K)
Calculated temp.
T (°C)
Measured
max. temp.
T (°C)
Calculated
temp.
T (°C)
Allowed
max. temp.
T (°C)
Remarks
1. Ambient
—
26.5
—
—
2. AC inlet
0.4
26.9
40.4
127.0
Added the 40°C
3. Wire between AC inlet and
Mains switch
8.9
35.4
48.9
161.0
Added the 40°C
4. Body of Main switch
1.5
28.0
41.5
127.0
Added the 40°C
26.0
52.5
66.0
169.5
Added the 40°C
6. PCB near U15(main board)
8.7
35.2
33.7
161.0
Added the 25°C
7. PCB near U14(main board)
10.1
36.6
35.1
161.0
Added the 25°C
19.0
45.5
59.0
203.5
Added the 40°C
6.7
33.2
31.7
-
Added the 25°C
10. Top enclosure
3.6
30.1
43.6
127.0
Added the 40°C
11. "Record" button
5.4
31.9
45.4
127.0
Added the 40°C
12. Applied part(DOP I)
0.5
27.0
40.5
41.0
Added the 40°C
13. Applied part(DOP II)
0.8
27.3
40.8
41.0
Added the 40°C
14. Applied part(UC)
0.7
27.2
40.7
41.0
Added the 40°C
15. Applied part(event Mark jack)
0.8
27.3
40.8
41.0
Added the 40°C
5. Body of Motor
8. Coil of Transformer
(AC/DC Adapter)
9. Body of enclosure
(AC/DC Adapter)
Comments
According to IEC 60601-1 clause 42.1, table Xa and Xb, 400C or 250C is added to the temperature rise
for determination of the final temperature T.