F9, F9 Express Fetal & Maternal Monitor Service Manual Disassembling the Monitor No. i ii (i) Connect to SpO2 Socket Bottom Interface Board (ii) 10.4.11 Replacing the ECG Module To remove the ECG module, 1) Separate the upper assembly and the lower assembly. 2) Disconnect all the wires on the ECG module. 3) Remove the two studs/screws securing the module. To replace the ECG module, Replace the ECG module in a reverse procedure. Refer to the following illustration when connecting wires. Ensure they are all correctly connected. (i) No. ( ii ) ( iii ) - 63 - Connect to i Bottom Interface Board ii TEMP Socket iii ECG Socket