Download pdf-file
Transcript
PULPO Doc: Issue Date Page VLT-MAN-ESO-xxxxx-xxxx 2010 21 of 65 HE,3,0 or use Menu Mode 1 in the front panel display. 10.2.2 Disable Alarms To avoid annoying buzzer and front panel warnings while the dewar variables (temperature, vacuum, LN2 exhaustion detector) are not stabilised , it is recommended to make a global alarm disabling with command AE,0,0. Take the opportunity to set the trip points for the temperature and vacuum alarms. In steady state operation, PULPO is capable of maintaining the reference temperature with less than 0.1K, so the trip point can be set 0.5 or 1K higher. TT,1,150.5 TT,2, 330 TT,3,330 .... VT,1e-4 // set point for sensor 1 is 150K and alarm trip point is at 150.5K // trip point for unused sensors set at very high value //vacuum alarm trip point at 1e10-4 mB 10.2.3 Heaters Interconnection Tell PULPO how the heaters are interconnected, using command DM,n. The ESO cryostats Generation 3 (VLT) have heater 1 and 2 working in parallel, and heater 3 is not used. So issue DM,1 or set PULPO front panel in Menu Mode 1 scroll till CABLING is displayed end press the EDIT key until 1=2 3 label is displayed. 10.2.4 Associate Pt100 sensor to Heaters To control the detector temperature through the PID controller programmed in software, we need to associate one Pt100 sensor to each heater. In this way, the software will read this sensor and compare its value with the reference value programmed either via RS232 link or through the PULPO front panel. This operation can only be done through the serial link. Use the command Control Sensor (CS,n,m), to tell PULPO that control sensor m is associated to heater n. If heater 1 and 2 are working in parallel, they must have the same temperature sensor associated, so give commands CS,1,1 and CS,2,1 if both are using sensor 1 as control sensor (Generation 3 cryostat). 10.2.5 Setting Heaters mode operation Set the heaters to work in PWM mode with 1 sec maximum duty cycle. For that issue commands: HM,1,1 HM,2,1 HM,3,1 10.2.6 Set PIDs Set Points Load the set point reference for the three heaters via Menu Mode 1 or commands SP,n,f. If there are heaters working in parallel (same reference Pt100 sensor) make sure to set the same set points for them. ESO, Karl-Schwarzschild-Str. 2, 85748 Garching bei München, Germany