Download Introduction - TEquipment.NET

Transcript
Model 6517B Electrometer Reference Manual
Section 14: SCPI Command Reference
:FRESh?
:CALCulate[1]:DATA:FRESh?
Description
Recalculate CALC 1
This query command is used to return a new (fresh) math result of CALC1. This
command will not request the same result twice. If a new reading is triggered, this
command will wait until the math result is available. When CALC1 is disabled,
there is no difference between SENS:DATA:FRES? and CALC:DATA:FRES?.
:IMMediate
:CALCulate[1]:IMMediate
Recalculate CALC 1
Perform calculation and read result
Query
:IMMediate?
Description
When you change the math calculation, the next instrument reading reflects the
new calculation (assuming CALC 1 math is enabled). If the instrument is not in a
continuous measurement mode (for example, waiting for a trigger), the displayed
reading will not update to the reflect the new calculation until the next reading
conversion occurs.
This action command allows you to re-process the current input data to reflect a
new calculation. For example, assume the instrument is in a non-continuous
measurement mode and requires a manual trigger to cause the next reading
conversion. Changing the calculation will not affect the currently displayed
reading. However, sending the :IMMediate command re-processes that
displayed reading to reflect the new calculation. Note that sending the
:IMMediate command does not initiate a reading conversion.
An alternate way to perform the calculation and read the result is by using the
query form of the command (:IMMediate?). When this command is sent, the
calculation is performed and the result is queried.
CALCulate2
This calculate subsystem is used to configure and control CALC 2 operations on readings stored
in the buffer.
:FORMat <name>
:CALCulate2:FORMat <name>
Specify CALC 2 format
Parameters
<name> = MEAN
SDEV
MAX
MIN
PKPK
NONE
Mean value of readings in buffer
Standard deviation of readings in buffer
Largest reading in buffer
Lowest reading in buffer
Peak-to-peak value of readings in buffer
No calculation
Query
:FORMat?
Description
This command is used to specify the format for the CALC 2 math calculation. The
calculation operations for CALC 2 use data stored in the buffer.
Query programmed math format
With NONE selected, no CALC 2 calculation will be performed. With one of the
other formats selected and CALC 2 enabled (see :STATe <b>), the calculation
is performed every time the :IMMediate or :IMMediate? command is
executed.
PKPK = MAX - MIN
6517B-901-01 Rev. B / June 2009
Return to Section Topics
14-27