Download Introduction - TEquipment.NET
Transcript
Model 6517B Electrometer Reference Manual Section 14: SCPI Command Reference :FRESh? :CALCulate[1]:DATA:FRESh? Description Recalculate CALC 1 This query command is used to return a new (fresh) math result of CALC1. This command will not request the same result twice. If a new reading is triggered, this command will wait until the math result is available. When CALC1 is disabled, there is no difference between SENS:DATA:FRES? and CALC:DATA:FRES?. :IMMediate :CALCulate[1]:IMMediate Recalculate CALC 1 Perform calculation and read result Query :IMMediate? Description When you change the math calculation, the next instrument reading reflects the new calculation (assuming CALC 1 math is enabled). If the instrument is not in a continuous measurement mode (for example, waiting for a trigger), the displayed reading will not update to the reflect the new calculation until the next reading conversion occurs. This action command allows you to re-process the current input data to reflect a new calculation. For example, assume the instrument is in a non-continuous measurement mode and requires a manual trigger to cause the next reading conversion. Changing the calculation will not affect the currently displayed reading. However, sending the :IMMediate command re-processes that displayed reading to reflect the new calculation. Note that sending the :IMMediate command does not initiate a reading conversion. An alternate way to perform the calculation and read the result is by using the query form of the command (:IMMediate?). When this command is sent, the calculation is performed and the result is queried. CALCulate2 This calculate subsystem is used to configure and control CALC 2 operations on readings stored in the buffer. :FORMat <name> :CALCulate2:FORMat <name> Specify CALC 2 format Parameters <name> = MEAN SDEV MAX MIN PKPK NONE Mean value of readings in buffer Standard deviation of readings in buffer Largest reading in buffer Lowest reading in buffer Peak-to-peak value of readings in buffer No calculation Query :FORMat? Description This command is used to specify the format for the CALC 2 math calculation. The calculation operations for CALC 2 use data stored in the buffer. Query programmed math format With NONE selected, no CALC 2 calculation will be performed. With one of the other formats selected and CALC 2 enabled (see :STATe <b>), the calculation is performed every time the :IMMediate or :IMMediate? command is executed. PKPK = MAX - MIN 6517B-901-01 Rev. B / June 2009 Return to Section Topics 14-27