8.2 Zero Event
A zero event determines the zero output value of the MD concentration sensor.
A zero event can be initiated two ways—AUTO or MANUAL. The method is set
from the SETUP:ZERO MODE screen.
The Zero event causes the follow state sequences:
•
Turn off the span calibrator
•
Reverse the air pump to purge the chamber with clean air.
•
Wait 45 seconds for the analog concentration output to stabilize
•
Sample for 15 seconds and calculate and store the mean zero value
•
Set the air pump forward to draw outside air into the chamber
•
Wait 30 seconds for the air pump to stabilize
•
Start sampling
8.2.1 Auto Zero Event
When set to AUTO mode a zero event occurs periodically at the period set from the
SETUP:ZERO MODE screen.
8.2.2 Manual Zero Event
When set to MANUAL mode the user can initiate a zero event from the operate
screen.
The MANUAL mode command is only available when the unit is on and sampling.
8.3
Span Event
A span event determines the span output value of the MD concentration sensor.
A span event can be initiated two ways—AUTO or MANUAL. The method is set
from the SETUP:SPAN MODE screen.
The Span event causes the follow state sequences:
•
Switch the MD engine to the 1,000 ug/m3 range
•
Turn on the span calibrator.
•
Wait 45 seconds for the analog concentration output to stabilize
•
Sample for 15 seconds and calculate and store the mean span value
•
Switch the MD engine to the user concentration range setting
•
Turn off the span calibrator
•
Set the air pump forward to draw outside air into the chamber
•
Wait 30 seconds for the air pump to stabilize