KIT COMPONENTS
Microwell plate
Negative Control
Cut-off Control
12x8 wells coated with EBV antigens in a
resealable foil pouch.
1 vial, 1.5 ml, IgA, IgG or IgM, human serum,
ready to use
Substrate Solution B
1 vial, 3 ml, IgA, IgG or IgM, human serum, ready
to use
1 vial, 1.5 ml, IgA, IgG or IgM, human serum,
ready to use
1 vial, 12 ml. Anti human IgA, IgG or IgM HRP
Conjugate, ready-to-use
1 vial, 0.750 ml. TMB (Tetramethylbenzidine) in
DMSO (Dimethylsulfoxide)
1 bottle, 15 ml, contains urea peroxide (corrosive).
Sample Diluent
2 bottles, 20 ml each.
Additive for Sample Diluent
2 vials, 2ml each
Wash Buffer
1 bottle, 20 ml. Contains NaCl and Tween 20.
(25x) Concentrate
1 bottle, 15 ml. 1 % v/v sulphuric acid, corrosive.
Ready to use
Positive Control
Enzyme Conjugate
Substrate Solution A
Stop Solution
MATERIALS REQUIRED BUT NOT PROVIDED
 Deionized or distilled water
 Graduated cylinders and beakers
 Wash trays
 Macropipettes capable of delivering 5 µl to 1000 µl.
 Multichannel Micropipette
 Stepper
 Incubator 37°C+/- 2°C (dry incubator, make sure to correctly seal the wells with
the adhesive foil to prevent evaporation which may lead to erroneous results.
 Microplate reader capable of reading absorbance values at 450 nm. If dual
wavelength microplate reader is available, the reference filter should be set at
600-690 nm.
 Automatic microplate washer capable of dispensing 300 µl.
Microplate reader and microplate washers are available from Abiyotek Company.