Troubleshooting and Customer Service
Symptom: Power is below specification
Possible Cause:
Corrective Action:
Diode current is set too low.
Verify the expected output power levels for the
Low, Medium and High settings.
Q-switch frequency is set
higher or lower than 1 kHz.
Power output depends on Q-switch frequency.
If operating the Evolution-30 at a different
Q-switch frequency than specified, contact
Spectra-Physics for the power specification at
that particular frequency.
Cavity optics are out of
alignment.
Adjusting the cavity mirrors slightly may optimize the output power. Read the detailed
instructions in Chapter 6, “Maintenance,”
before attempting to adjust the mirrors.
LBO crystal temperature has
drifted.
At low pump power, check the output power
across the 3-degree tuning range of the LBO
crystal for the highest output power.
Algae growth is present in the
cooling water.
Follow the maintenance procedure in Chapter 6.
Diode laser power is low.
After several thousand hours of operation, the
optical power of the diode lasers decreases
and the current to the diodes must be
increased to compensate. Contact your service representative.
Symptom: Laser does not emit Q-switched or CW output
Possible Cause:
Corrective Action:
External Q-switch input is not
valid.
If an external Q-switch signal is being used,
verify the signal is at least 2 V into a 50-ohm
load and at least 2.5 µs wide; also verify the
frequency is the desired value (between 1–
10 kHz).
QSW MODE is set to CW RF
OFF.
Change the setting of QSW MODE to INTERNAL.
QSW BNC cable is removed.
If either QSW cable has been removed, the
Evolution-30 will report a VSWR error and
change the QSW MODE to CW RF OFF. Verify both BNC cables are attached at both the
power supply and laser head.
7-3