Download Manual - Sin
Transcript
Important: general instructions for remote control inputs All remote control inputs can be wired in two ways: Logic A logic voltage between 5V TTL and 24V industrial logic may be wired to the input pin. The pump is configured to operate without modification anywhere in this range. Any of the 0V pins is connected to the control device’s 0V to make the circuit. One 0V pin is identified in the wiring diagram, but any may be used. Low is equivalent to 0V. High is equivalent to 5V→24V. Switch A remote switch may be connected between the input pin and any positive voltage supply from the pump. One 5V supply pin is identified in the wiring diagram but there are several which will do, either 5V or other positive voltage. However, do not use pin 21 on the lower Dconnector or pin 20 on the upper D-connector. 21.4 Run/stop input Remote run/stop: connect a remote switch between pins 7 and 19 of the lower Dconnector. Alternatively a logic input may be applied to pin 7 of the lower D-connector, ground to pin 18. High input stops the pump, low input runs the pump. With no connection or with the switch open, the pump will default to running. To change or set the sense of the run/stop input, see 16.5 Remote stop in the Setup menu. 21.5 Direction input To enable remote direction control and disable the DIRECTION (SHIFT, 1) key on the keypad, link pins 6 and 18 of the lower D-connector. Connect a remote switch between pins 5 and 19 of the lower D-connector. Open switch for clockwise rotation, Watson-Marlow 620UN, 620U, 620SN, 620S User Manual 64