Operating manual - English
UNOCODE 399
ATTENTION: do not touch the X and Y axes carriages and
press START.
11) press START
Pos. sensors.
ICX = ON
ICY = ON
See operating manual.
[START]
START
- the ‘machine’s zero’ settings are detected by electric contact
- when this operation is completed, the
machine’s display will show the measured quotes
ATTENTION: if the STOP key is pressed, the new settings
will be lost. If so, only the previous setting
values will remain valid.
12) press ENTER to save the settings.
13) remove the (Z1) and (Z2) templates.
14) proceed with setting the reading unit.
READING UNIT
Zero detection
in progress !
Measurements read:
X = 4444
Y = 3333
Save data ?
No = STOP
Yes = ENTER
ENTER
1 - Cutting unit
2 - Reading unit
1) Select ‘Reading unit’ #2.
2) press ENTER.
3) remove the R100 clamp (optical reader
side) from its support.
4) fi the (Z2) template and the clamp’s
bushing securing it with the clamp’s knob
(fig. 64).
5) lower the protective shield and press
START.
- the ‘machine’s zero’ settings are detected by the optical reader.
- when this operation is completed, the
machine’s display will show the measured quotes.
ATTENTION: if the STOP key is pressed, the new settings
will be lost. If so, only the previous setting
values will remain valid.
ENTER
Install the Z2 template.
See operating manual.
[START]
START
Zero detection
in progress !
Measurements read:
X = 4444
Y = 3333
Save data ?
No = STOP
Yes = ENTER
ENTER
6) press ENTER
7) remove the (Z2) template.
- the new setting has been made.
clamp’s knob
clamp’s bushing
Z2
Fig. 64