COPS Readout Board
User’s manual
4. Response to the entered TT or CD commands has a distortion.
Check a configuration of the serial port used for communication.
Check the used serial port with the reference readout board.
Check all cable connections.
If all external staff is valid, follow point 3 of this chapter.
5. Operation with some board numbers has no response; other board numbers are
valid.
Find shortened conductor or component and eliminate defect.
6. The DAC cannot be loaded.
Input the AP 1 command (+9V and +5V turning on).
Input command SD 1000 1 (loading of the DAC in loop).
Check communication of the DAC and DSP with oscilloscope.
Eliminate found defects.
7. Data received with disconnected CCD sensor exceed 20 hexadecimal.
Check output voltage of the DAC; it must be zero. Otherwise, see point 6 of
this chapter.
Check values of all resistors in the front-end circuitry.
Input command CC 3 1 (CCD data conversion in loop).
Check +9V and +5V voltages; they must be turn on before a conversion and
turn off after it.
Check all output signals produced by the timer to control the CCD sensor
(0RS, 0ROG, 0CLK), the front-end (BASE, PEAK), the ADC (SCLK, ADCS,
ADCD), and their timing. Follow these signals from the timer CPLD to
targets.
Check communication between the ADC and DSP.
Eliminate found defects and check the readout board once more.
8. Data received with the connected reference sensor has no peak
Check a cable connection between the CCD sensor and the readout board.
Follow the CCD sensor output signals from the sensor connector P1 to the
ADC.
Eliminate found defects and check the readout board once more.
Page 29 of 43