Configuration and structure of measuring system Yamatake Corporation 1-1 : Loop configuration 1-1-1 : Measuring system Introduction This instrument is an electromagnetic flowmeter and outputs analog and pulse or contact output. It measures the flow rate of a conductive fluid flowing in the sensor and outputs a signal according to the set range. Concept of flow rate measurement by this instrument The concept of a flow rate measuring system using this instrument is shown. Output signal Analog output Pulse output Contact output Fluid Figure 1-1 Conceptual drawing of measuring system (integral type) Analog output If instantaneous flow rate values are to be output as an analog variable to control equipment, configure the system to deliver an analog output. Pulse output As a pulse output, the open-collector output is available. Pulse frequencies up to 200 Hz can be delivered. The pulse output can be delivered simultaneously with an analog output. Contact output Instead of the pulse output, a contact output is available using the open-collector output. The contact output can be delivered simultaneously with an analog output. 1-2 Model MCB10A