C-Link Protocol Commands
Measurements
detector base temp
This command reports the current detector base temperature, in degrees C.
The following example reports that the current detector base temperature is
174.9 °C.
Send:
Receive:
detector base temp
detector base temp 174.9 deg C
filter temp
This command reports the current sample filter temperature, in degrees C.
The following example reports that the current filter temperature is 175.1
°C.
Send:
Receive:
filter temp
filter temp 175.1 deg C
flame temp
This command reports the current FID flame temperature, in degrees C.
The following example reports that the current flame temperature is 221.0
°C.
Send:
Receive:
flame temp
flame temp 221.0 deg C
fid signal
This command reports the average raw FID signal, in Hz. The following
example reports that the fid signal is 4577.
Send:
Receive:
fid signal
fid signal 4577
internal temp
This command reports the current internal instrument temperature. The
following example reports that the internal temperature is 25.3 °C.
Send:
Receive:
internal temp
internal temp 025.3 deg C
c1 value
nm value
thc
thc value
These commands report the measured methane, non-methane, and total
hydrocarbon concentrations. The following example shows that the THC
concentration is 40 ppm.