2.4 Adjusting the Chart Speed
Relationships between chart speeds and their code numbers are shown below.
Chart Speed (unit: mm/h)
10
15
20
25
30
40
50
60
75
80
90
100
120
150
200
300
600
1200
1500
3000
6000
The data for a dot model (10 mm/h to 1200 mm/h) are shown within the bold border.
Setting Procedures
Press the DISP key for three seconds to enter the Regular
ESC
DISP
Setting mode. Then press the ENT key.
3 sec
..
.
CH UP
Select [Chart] by pressing the
key.
key, and press the ENT
Using the
key, select one of the two kinds of chart speed,
the No.1 and No.2 chart speeds. Then press the ENT key.
The No.2 chart speed identifies the chart speed which takes
effect in case the speed is changed by remote control
CH UP
(optional function).
For models without the remote control function, the setting
of the No.2 chart speed is invalid.
Select a numeric code referring to the table above using the
key and press the ENT key.
The setting ends with the [Chart] indication.
When you want to continue to set another chart speed ,
press the ENT key. In order to return to the operation mode,
press the DISP key for three seconds.
2-10