2 – 26 2.6.2 Remote Operation Query Errors Error Code 30 Meaning Lost Data Data in the output buffer was lost. This occurs if the output buffer overflows, or if a communications error occurs and data in output buffer is discarded. 32 No Listener This is a communications error that occurs if the SR470 is addressed to talk on the GPIB bus, but there are no listeners. The SR470 discards any pending output. 2.6.3 Device Dependent Errors Error Code 40 Meaning Failed ROM Check The ROM checksum failed. The firmware code is likely corrupted. 2.6.4 Parsing Errors Error Code 110 Meaning Illegal Command The command syntax used was illegal. A command is normally a sequence of four letters, or a ‘*’ followed by three letters. 111 Undefined Command The specified command does not exist. 112 Illegal Query The specified command does not permit queries 113 Illegal Set The specified command can only be queried. (Continues) SR470 Shutter Controller