794380D / 0914 4. FIRMWARE COMMANDS AND RESPONSES ENABLE_GAIN_STABILIZATION Enables the stabilization of the gain peak. See also DISABLE_GAIN_STABILIZATION, SHOW_GAIN_STABILIZATION, and INITIALIZE_GAIN_STABILIZATION. ENABLE_HV Turns on the high-voltage enable output of the DSPEC Pro. See also DISABLE_HV and SHOW_HV. ENABLE_HV_DIM Enables the DSPEC Pro to work with a standard DIM (e.g. DIM-POSNAI, DIM-POSGE, or DIM-NEGGE, DIM-296). This command causes the DSPEC Pro to poll the DIM and return state-of-health information to a controlling program (see SHOW_HV_HEALTH command). ENABLE_LFR Sets the DSPEC Pro to LFR mode. See also DISABLE_LFR and SHOW_LFR. ENABLE_NAI Switches the stabilizers to NaI mode. The stabilizers use the “fine” adjustment in NaI mode and the “very fine” adjustment in Ge mode. See DISABLE_NAI. ENABLE_OPTI_NEED Enables the reporting of an error when the DSPEC Pro needs optimizing. ENABLE_OVERFLOW_PRESET Enables the overflow preset. Channels that receive a count when they contain 2147483647 counts, the maximum number of counts, will stop the acquisition if the overflow preset is enabled. The channel that caused the preset to complete will contain 2147483647 counts. See also DISABLE_OVERFLOW_PRESET and SHOW_OVERFLOW_PRESET commands. ENABLE_PZ_AUTOMATIC Enables the automatic pole zero mode. See also DISABLE_PZ_AUTO and SHOW_PZ_ AUTO. ENABLE_PZ_NEED Enables the reporting of an error when the DSPEC Pro needs pole zeroing. ENABLE_REMOTE No function in the DSPEC Pro; included for backward compatibility. 59