broadband segment and a CW segment. Set the broadband frequency list segment range
at least as wide as the operating range of the device.
Set the CW frequency segment to the frequency you intend to use for the data acquisition,
the same frequency you set in the model variable table at the beginning of this procedure.
This frequency should be above 200 MHz to eliminate the effects of test set coupler roll-off
at low frequencies, and low enough to eliminate bond inductances and other parasitic
inductance considerations.
Based on the broadband cal, use a calibrated measurement point equal to the start
frequency plus an integer multiple of the step size. An example frequency might be 600
MHz.
200 MHz (start freq)+1·400 MHz (step size)= 600 MHz
Frequency List Calibration for an Agilent 8510­Based System
If you are using an Agilent 8510-based system, follow the steps below.
1. On the network analyzer press PRESET.
2. To define a linear frequency list, on the network analyzer press STIMULUS > MENU
> MORE > EDIT LIST > ADD. A default frequency list segment is presented.
3. To change the segment to your chosen frequency range press SEGMENT: START,
and use the numerical keypad to set the start frequency, ending with one of the
terminator keys (such as M/μ) at the right of the keypad. Similarly, press STOP and
set the stop frequency.
Note
Use the SEGMENT: START and STOP softkeys, not the START and STOP hardkeys in the STIMULUS
area of the front panel.
4. Press SEGMENT: NUMBER of POINTS, and enter the number of frequency points to
be measured across the range. Keep in mind that the fewer the number of points, the
less time the measurements take.
5. Press DONE to return to the frequency list menu. The segment you just defined is
listed on the screen.
6. Press DONE to return to the main stimulus menu. Press FREQUENCY LIST to
activate the frequency list mode.
7. Press PRIOR MENU > MORE > CONTINUAL to set a continual stimulus sweep.
8. Press PRIOR MENU > POWER MENU, and set the desired power level. In setting
the source power and attenuation, take care that the power level will not be
excessive at the device input. Also consider the gain of the device, and set a power
level that will not saturate the input port samplers of the analyzer. If the power level
at the sampler goes above ≈-8 dBm, an *IF OVERLOAD* error message is displayed
and you will need to reduce the source output power. The default network analyzer
power level is 0 dBm.
9. For a device with power drop-off at higher frequencies, you may wish to set a power
slope using the stimulus menus. An appropriate power slope would be in the region
of 2-3 dB/GHz.
10. Press RESPONSE MENU > AVERAGING ON/restart, and enter an averaging factor
high enough to reduce trace noise and increase dynamic range as appropriate for
your device measurements. Although the default is 256, you may wish to use an
averaging factor as low as 16 to reduce overall measurement time.
11. If the cal kit constants for your calibration kit are not loaded into the network
analyzer, load them from disc now by pressing DISC > LOAD and specifying the
data type and file name.
12. If you wish to modify one of the internal cal kit definitions, do so now.
13. Press CAL > CAL1 or CAL2 (depending on your cal kit), and perform a two-port
calibration, measuring each of the standard devices in turn and pressing the softkeys
as each measurement is complete.
14. At the end of the calibration sequence, the cal set numbers are listed in the softkey
menu. Press CAL SET #1 to store the calibration in cal set #1. (Or use another
available cal set.)
CW Calibration Subset for an Agilent 8510­Based System
1. To create a CW frequency subset of the broadband frequency list calibration, press
CAL > CORRECTION ON > CAL SET 1 (this turns on the broadband cal you just
completed), then MORE > MODIFY CAL SET > FREQUENCY SUBSET. Use the
SUBSET: START and SUBSET: STOP softkeys in the SUBSET menu ( NOT the front
panel keys) to set both start and stop to the same frequency.
2. Press CREATE & SAVE, and store this calibration in cal set #2 or another available
cal set.
3. Note the numbers of the cal sets where you save the calibrations: you will need to
list one of these numbers as part of the instrument state for each of the network
analyzer measurements.
Note
For more detailed information, refer to the Agilent 8510 Operating and Programming Manual.
Frequency List Calibration for an Agilent 8753­Based System
If you are using an Agilent 8753-based system, follow the steps below.
330