Appendix A: Disinfectants
Disinfect the IOD once per week using the disinfectant listed in the table below. Wet a cotton
ball or cotton swab and gently wipe the surfaces of the IOD. DO NOT SCRUB! DO NOT USE
CLEANERS! Thoroughly rinse the IOD with tap water to remove any residual disinfectant and
allow it to air-dry. DO NOT use a hair dryer or the like to force dry the IOD.
The disinfectant listed below have been tested and shown to be compatible with the IOD. Follow
the manufacturer’s directions regarding safe use and disposal for the disinfectant.
Please visit our website for the most recent list of approved disinfectants. Disinfectants are
added to the list as they are tested and shown to be compatible with the IOD.
Disinfectant
70% isopropyl alcohol
(rubbing alcohol)
Source
Over the Counter (local pharmacy)
Wicab, Inc.
DR-000807 Rev H, March 2013
- 21 -
®
BrainPort V100-T Device, V100
Trainer Manual