Transcript
Preparing for a Run—ABI PRISM 377, ABI PRISM 377 with XL Upgrade, and
ABI 373 with XL Upgrade
The specific equipment and reagents needed for a microsatellite analysis run are
listed in “Before You Begin” on page 8-3.
Gel Preparation Note
Appendix A contains protocols for preparing both Long Ranger and polyacrylamide
denaturing gels.
For the ABI PRISM 377 and ABI PRISM 377 with XL Upgrade
Applied Biosystems recommends using a 5% Long Ranger/6 M urea gel.
Alternatively, you can use a 4.25% polyacrylamide/6 M urea gel.
Use:
♦
36-cm well-to-read plates
♦
0.2-mm spacers
♦
Square- or shark’s tooth comb
For the ABI 373 with XL Upgrade
Applied Biosystems recommends using a 5% Long Ranger/6 M urea gel.
Alternatively, you can use a 6% polyacrylamide/8.3 M urea gel.
Use:
♦
12-, 24-, or 34-cm well-to-read plates
For most applications, you can obtain acceptable results using a 12-cm gel. If time
is not limiting, longer plates will give better resolution and greater precision.
♦
0.4-mm spacers
♦
Square- or shark’s tooth comb
Instrument Setup Refer to the ABI PRISM 377 DNA Sequencer User’s Manual, the 373 DNA Sequencing
System User’s Manual, or the 373 DNA Sequencer with XL Upgrade User’s Manual for
the general procedure.
Table 8-2 Run Window Settings for the ABI PRISM 377 and 377XL
Data Field
Setting
Plate Check Module
Plate Check (A, C, D, or F)
PreRun Module
PR GS 36A-2400 (or 36C, D, or F)
Run Module
GS 36A-2400 (or 36C, D, or F)
Collect Time
2 hours
Note
To see all product fragments you might
need to increase the run time to 2.5 hours.
8-8 Microsatellite Analysis
Sample Sheet
The name of the sample sheet that you created for
this run
Well-to-Read distance
36 cm
Lanes
As appropriate for the comb used