TRIAXIAL TESTING
© 2013 Geocomp Corp.
*Note: There is no particular requirement for the quality of the chamber water as long
as the water is clean. Water that is drinkable satisfies this requirement.
II-2. Q: Do I need to use de-aerated water to fill the FlowTrac-II units; if so, how do I
obtain de-aerated water?
A: Yes, it is important to use de-aerated water in the FlowTrac-II pumps. This will
minimize the water air content thereby speeding up the back saturation phase. To prepare
de-aerated water, boil water while applying a vacuum or use a device such as a Nold
DeAerator, which combines propeller cavitations with a vacuum to de-aerate the water
rapidly.
II-3. Q: Could the Node ID# for either one of the FlowTrac-II units or the LoadTrac-II unit
change by itself? What possible reasons could you think to have prompted the change?
A: The imbedded controller default Node IDs for LoadTrac-II and FlowTrac-II are 65 and 66
respectively. When a second FlowTrac-II unit is supplied with a Triaxial system, its ID is set at
the factory as 67. The node IDs cannot change by themselves. However, if you reboot the
second FlowTrac-II unit (the one used for controlling the sample pressure) and by mistake
press the ESC key on the keypad at the beginning of the boot process, the control program
reverts back to the default value of 66. Similarly, if for some reason you are using different
Node IDs than those specified above (because you are controlling several systems with one
computer), pressing the ESC key at the beginning of a reboot will change the ID for the unit
being rebooted back to its imbedded controller default value.
If there is a possibility that someone pressed the ESC key at the start of the boot process
for one or more units in your system, go through the following steps to reset default values
for the control programs.
1. Turn off the LoadTrac-II unit and the two FlowTrac-II units.
2. Wait about 10 sec.
3. As you turn each back on, immediately press the ESC key on the keypad to reset
everything back to the imbedded controller default values.
4. Now go to your LoadTrac-II and FlowTrac-II hardware manuals. In Section 2.1.4 of
each manual (Setup section) you will find a table of default settings for your system.
Got through each LCD menu and be sure the settings on the menu agree with those in
the table. If a setting does not agree, change it to the one in the manual.
5. If the Node ID numbers on your units were changed (because you are controlling more
than one system with your computer), set them to the values they were changed to.
(The factory-setting default values are 65 for LoadTrac-II, 66 for the FlowTrac-II unit
controlling cell pressure and 67 for the FlowTrac-II unit controlling sample pressure.)
II-4. Q: We noticed that a lot of the clay in our specimen has been washed into the sample
flow pump. We think that the flow pump might now contain soil deposits after some time
of use. What is your opinion? Do you have a similar experience? Does the washing of the
fine materials into the flow pump affect the readings of the pressure transducer?