Download Report 2 - CSA Group

Transcript
S ECTION 5
MAINTENANCE AND CALIBRATION PROCEDURES
10. The Enter Filter Weight screen will display (Figure 5-85). Select the
Filter Weight: button. The keypad will display. Enter the weight of the
pre-weighed filter into the system and press the Enter button to save
the value and exit the keypad. Select the Next > button.
Figure 5-85.
Enter Filter Weight screen.
THERMO FISHER SCIENTIFIC
OPERATING GUIDE, TEOM 1405-DF
5-69