6
Preventive Maintenance and Repair
Detector
3 Unscrew, disconnect and replace the lamp. Insert, fix and 4 Replace the heater assembly.
reconnect the lamp.
Next Steps:
5 Replace the front cover.
6 Reset the lamp counter as described in the Utilities software documentation.
7 Turn the lamp ON.
8 Give the lamp more than 10 minutes to warm-up.
9 Perform Wavelength Calibration to check the correct positioning of the lamp.