Download Manual - ETS
Transcript
HIn3637 Manual Page — 9 assistance. Controls, Indicators and Connectors The instrumentation is housed in a rugged aluminum chassis. The entire unit is shipped in a convenient carrying case. The protective foam liner inside the carrying case has cutouts which accommodate the meter, the detachable probe and the optional accessories offered for the meter. The probe consists of a sensing head containing three orthogonal coils oriented to pick up the X, Y, and Z components of the field. The detachable probe plugs into the multi-pin connector mounted on top of the chassis. The color coded label surrounding the probe connector will match the color of the band on the probe cover. The top of the chassis also contains a ¼ inch stereo phone jack. The jack is used to connect the battery charger and also provide access to the recorder output signal. The rotary selector switch on the HIn3637 front panel controls power to the instrument and selects one of five measurement ranges. The scales are arranged so that when turned to the first position—400 :T/4 G (or 300 Amps/meter if the optional kit is installed)—the meter is measuring on its least sensitive scale, i.e., highest range. As the switch is rotated clockwise, the sensitivity of the meter increases. As a general rule, readings will be most accurate when the meter movement shows maximum deflection without lighting the Over Range indicator. There are two indicators on the front panel. A "Low Battery" LED illuminates when the battery voltage is too low to provide accurate readings. At this point, testing should be discontinued and the battery recharged or replaced. The "Over Range" LED illuminates when the measured flux density exceeds full scale for the selected range.