RDC-141, User's Manual
(7)
(8)
(9)
Regulate the gas flow rate to 50 L/h by turning the valve on the rotameter.
When the furnace temperature has stabilised at 1000C, an alarm will
sound.
Calibrate the apparatus as outlined in #4.5.
4.5 Calibration
The calibration procedure should be performed once a week and after any maintenance of the apparatus (reactor tube or thermocouple replacement etc).
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(12)
(13)
(14)
(15)
Turn on the main switch on the front panel.
Set the calibration dial on the side of the control box to 100 minutes.
Set the temperature controller on the front panel to 1000C.
Insert the empty inner quartz tube and fix the cover with the clamp.
Activate the furnace and the gas by pushing the "FURNACE ON" button.
The lamps "Furnace on"
"CO2 on"
"Sample out" are alight.
Regulate the gas pressure to 2 bar by turning the valve on the control panel
directly below the pressure gauge.
Regulate the gas flow rate to 50 Nl/h by turning the valve on the rotameter.
When the furnace temperature has stabilised at 1000C, an alarm will
sound.
After the standby temperature is reached, the weighed standard sample
(5±0.001g) can be placed in the inner quartz tube and the cover refixed
with the clamp. The calibration standard is provided by R&D Carbon Ltd.
who have carried out a round robin involving at 10 apparatus measuring at
least one coke sample so that the average value can be statistically guaranteed.
Press the "SAMPLE IN" button to start the measurement cylce.
The lamps "Furnace on"
"CO2 on"
"Sample in"
are alight.
After 100 minutes the furnace will switch off automatically and 45 minutes
later an alarm will sound.
Take the internal reaction tube out and place it in the holder to cool for about
10 minutes.
Reset the furnace and the gas by pushing the "FURNACE ON" button.
None of the lamps will be alighted.
Weigh the remaining sample to a precision of ±0.001 g.
Calculate the reactivity in % weight loss as follows:
func R sub {CO sub 2} ~ = ~ {initial weight ~ - ~ final weight} over {initial weight}
~CDOT ~ 100 over 1 %
(16) Set the new calculated time on the calibration dial according to the equation: