Download Power Viewer Plus - Software Manual

Transcript
R&S Power Viewer Plus
Uncertainty Calculation
Data sheets often express the impedance matching of a device as a
standing wave ratio (SWR). The relationship between the SWR and the
reflection coefficient is expressed by the following equations:
s=
1r L
1−r L
r L=
s−1
s1
The example below demonstrates the influence of mismatch caused by
a signal source that is directly connected to a power sensor:
Load:
Source:
R&S NRPZ11
R&S SMBV100A
SWR = 1.2
SWR = 1.6
rL = 0.09
rG = 0.23
UdB = 2 · 0.707 · ± 8.7 dB · 0.09 · 0.23 = ± 0.25 dB
Manual
20