TECHNICAL INFORMATIONS
• Control unit
Dimensions: L=550 mm, l=450 mm, h=270 mm
• Gas blower
Diameter=50 mm
Height=80 mm
Weight=680 g
Materials: inox, aluminium, alumina and alumina fibers
Power : 220 V - 1000 W (fuse 10A)
Gas supply : 6 bars, 600 l/h maximum flow
Water cooling supply : maximum 6 bars, water flow 30 l/h minimum, inlet
temperature 25 °C.
Features and limits
• Maximum temperature : 1000 °C with a gas flow set to 500 l/h (air).
• Repeatibility in the isotherm zone : ≤ ± 2°C
• Stability during temperature dwell : ± 0.5°C
• The sample temperature corresponding to the setpoint temperature
must be defined by a calibration. The calibration must take into account:
- the emissivity of the sample
- the geometry
- the position in the flux
- the gas flow