Use Case
For this sample we will be setting up the use case below.
Settings:
Acqusition Mode = Continuous Frame
Acqusition Start Trigger Mode = Off
Frame Start Trigger Mode = ON
Frame Start Trigger Source = Line 1
Frame Start Trigger Activation = Falling Edge
Lines Per Frame (Height) = 3
Line Start Trigger Mode = On
Line Start Trigger Source = Frequency converter (from Shaft Encoder Module Output)
Line Start Trigger Activation = Rising Edge
Acquisition start
command
Acquisition start
command
Acquisition start
trigger signal
Frame start
trigger signal
Line start
trigger signal
Time
= Trigger signal internally generated by the camera
= Trigger signal applied by the user
= Camera is waiting for an acquisition start trigger signal
= Camera is waiting for a frame start trigger signal
= Camera is waiting for a line start trigger signal
= Line exposure and readout
= Line start trigger signal is ignored because the
camera is waiting for a frame start trigger signal
= Frame transmitted