Instruction Manual DVH / DVE Heinrichs Messtechnik Chapter 2 Installation Installation Overview Heinrichs’ DVH /DVE Vortex Flow Meter installations are simple and straightforward. Both the Series DVH In-Line and Series DVE Insertion type flow meter installations are covered in this chapter. After reviewing the installation requirements given below, see page 19 for Series DVH installation instructions. See page 22 for Series DVE installation instructions. Wiring instructions begin on page 36. Flow Meter Installation Requirements Before installing the flow meter, verify the installation site allows for these considerations: Line pressure and temperature will not exceed the flow meter rating. Warning! Consult the flow meter nameplate for specific flow meter approvals before any hazardous location installation. The location meets the required minimum number of pipe diameters upstream and downstream of the sensor head as illustrated in Figure 2-1. Safe and convenient access with adequate overhead clearance for maintenance purposes. Verify that the cable entry into the instrument meets the specific standard required for hazardous area installations. For remote installations, verify the supplied cable length is sufficient to connect the flow meter sensor to the remote electronics. Also, before installation check your flow system for anomalies such as: Leaks, valves or restrictions in the flow path that could create disturbances in the flow profile that might cause unexpected flow rate indications Chapter-Page 2- 17