Confirm settings with D [yes]. If settings are not correct press B [no] and repeat entry of
temperature settings.
After you have confirmed the temperature settings, the following screen is displayed. Here
you can enter all parameters for electrophoresis.
7.1.5
Enter electrophoresis parameters
1:L0: 30.0 L10: 70.0
time:
El: 0V 500mA 50W
A? B V*h C quit D
Enter time for electrophoresis and accept with D [enter].
Note: There is a convention on how time settings are entered in all BIOMETRA instruments:
hours
minutes
seconds
If you enter a number without “dot” this value will be interpreted as seconds (“300” => 5
minutes). To program minutes enter a “ ” after the number of minutes. To enter hours enter
after the number. You can also enter any combination of hours, minutes and seconds.
Example: for 1 hour, 30 minutes, 20 seconds enter 1 30 20.
The time values will be displayed in the following format: 00 m 00s
Accept time setting with [D enter]
1:L0: 30.0 L10: 70.0
time: 10m 0s
El: 0V 500mA 50W
A? Bdelete C quit D enter
Enter Voltage and accept with D [enter]
1:L0: 30.0 L10: 70.0
time: 10m 0s
El: 300V 500mA 50W
A? Bdelete C quit D enter
Note: The values for amperage [mA] and wattage [W] are set to maximum level as default. If
you enter lower values, these parameters may become limiting during electrophoresis.
Accept default settings for voltage and wattage with D [enter] or enter different values.
Note: We recommend to control electrophoresis by constant voltage rather than by constant
amperage (set [mA] to maximum value, respectively accept default value).
Note: Each program can consist of several steps. Thus it is possible to program complex
protocols including a pre-run, a pause for handling of the gel and the main run..
In the following screen you can program a second step for your protocol.
2:L0:
L10: