Download EMA10 - Megacon

Transcript
EMA10 - USER MANUAL
2.2)
MEASURING METHOD AND ACCURACY
Measuring range
30-500Hz.
Measuring method
64 sampling per period for V1 and A1, V2 and A2, V3 and A3.
Measuring interval 0,1 second.
Instrument accuracy
Model EMA10
Voltage
< 0.5 %
Current
< 0.5 %
Power
<1%
Energy
<1%
Power Factor
<1%
Standard
EN 61036
Model EMA10-05
< 0.25 %
< 0.25 %
< 0.5 %
< 0.5 %
< 0.5 %
EN60687
Sampling frequency
45 Hz=2.280 or at 60 Hz = 3,88kHz
Zero self-regulation
Offset
0,1 second.
RTC - REAL TIME CLOCK
Accuracy: 5 PPM, standard CEI-EN 61038
2.3)
PROGRAMMABLE PARAMETERS (SETUP SECTION)
VT and CT ratio.
Mode, insertion type (4 wires, 3 wires, Aron).
Integration time of Average parameters.
Sampling frequency.
Address or logical number of equipment.
Date and time.
Time-bands of power consumption in different periods.
Storage section (Min/max, Harmonics, average power and Sampling values).
All parameters concerning the input/output section (serial port, analog output, digital input and output).
Preset energy counters.
-8-