Download Model DY Vortex Flowmeter Model DYA Vortex Flow
Transcript
2. GENERAL DESCRIPTION Description Figure Valve positon (T-type piping exist): When pulsation causes by a T-type piping exist, install the valve on the upstream of the flowmeter. Example:As shown in the figure, when the valve V1 is turned off, the fluid flow throught B as to meter A the flow is zero. But due to the pulsating pressure is detected, the meter is zero point become fluctuating. To avoid this, change the valve V1 location to V1'. • In case of the Reduced Bore type, moisture may be remained upstream of the flowmeter. Drain it appropriately. Relocating digitalYEWFLO Valve (Off) Flow B V1 V1’ A Pressure and Temperature Taps: Pressure tap digitalYEWFLO Pressure tap outlet: install this tap between 2D and 7D on the downstream side of a flowmeter. Temperature tap outlet: install this on the downstream side 1D to 2D away from a pressure tap. Temperature tap Upstream Flow downstream 2 to 7D 1 to 2D digitalYEWFLO Mounting Gasket: Avoid mounting gaskets which protrude into the pipe line. This may cause inaccurate readings. Use the gaskets with bolt holes, even if digitalYEWFLO is the wafer type. When using a spiral gasket(without bolt holes), confirm the size with the gasket -manufacturer, as standard items may not be used for certain flange ratings. Pipeline Flange Pipeline No good digitalYEWFLO Heat-Insulation: Bracket When an integral-type flowmeter or a remote type detector is installed and the pipe carrying higt-temperature fluids is heat-insulated, do not wrap adiabatic materials around the installation bracket of the converter. • Refer to “2.5 Cryogenic and High Process Temperature Version Insulation” and install it rightly. Heat-Insulator Flushing of the pipe line: Flush and clean scale, incrustation and sludge on the inside of pipe for newly installed pipe line and repaired pipe line before the operation. For flushing, the flow should flow through bypass-piping to avoid damaging the flowmeter. If there is no bypass-piping, install short pipe instead of the flowmeter. digitalYEWFLO Short pipe F020102-2.EPS 2-3 IM 01R06A00-01E-E 3rd edition, Jan. 2007