2.1.2
Radiation Output Reproducibility
This test monitors the effect of the exposure parameters (tube voltage and
mAs) on the radiation output. Comparison should be made with the baseline
values established at commissioning.
Example: Measure at a range of tube voltages e.g. 70, 80, 90kVp at a range of
typical clinical mAs settings.
Note: Many CBCT units do not allow a manual selection of tube voltages and
mAs.
For these units, the above exposures should be made at the
automatically selected exposure settings.
2.2
Tube potential
The voltage applied to the X-ray tube determines the energy of the X-ray
photons and is a major factor in determining the contrast in the image.
Assessment of the tube potential ensures that the delivered kVp is close to that
set on the unit by the operator. Poor agreement between the two would affect
clinical image quality, equipment radiation output and patient dose.
2.2.1
kV accuracy
The kVp should be measured directly using a kV divider device at intervals of
10kVp across the full range the unit is capable of producing.
2.2.2
kV repeatability
The consistency of the tube potential should be monitored by repeating five
measurements at at least two clinically relevant kVp values, where possible.
2.2.3
kV reproducibility
The reproducibility of the tube potential over time should be monitored by
comparing the measured results for kVps at intervals of 10kV across the full
range the unit can produce with those established as baseline values at
commissioning.
2.3
Filtration
The filtration of an X-ray tube absorbs the low energy photons that do not
contribute to the image formation but do contribute to patient skin dose. Having
adequate filtration is essential to ensure that patient dose is controlled. The
total filtration should be marked on the X-ray tube housing.
Total filtration can be estimated by measuring the Half-Value Layer (HVL). The
HVL is the thickness of the absorber required to reduce the intensity of the
incident X-ray beam by half. The HVL is an estimate of the penetrating power of
the X-ray beam which means that the higher the HVL the more penetrating the
X-ray beam is.