Preparing the Reaction Plate
Standard vs. Fast Plates
STANDARD
Preparing the Reaction Plate
Standard vs. Fast Plates
IMPORTANT! Make sure that you use the standard Optical 96-Well Plate on the 7500
Real-Time PCR system. Optical 96-Well Fast Plates will not fit into the standard block
correctly and will result in loss of data.
Fast Plates
Standard Plates
Standard plate has
notch at top-right
corner by A12
Fast
plate has
notch at
top-left
corner by
A1
30-µL
maximum
reaction
volume
GR2386
7900
HS96 well plate comparison
100-µL
maximum
reaction
volume
1. Label the reaction plates, ensuring that you include an endogenous control for each
sample type (for example, each tissue in a study comparing multiple tissues). If
samples are spread across multiple plates, each plate must have an endogenous
control. Additionally, every plate must include an endogenous control for every
sample type on the plate.
4
2. Into each well of the reaction plate, add 50 µL of the appropriate PCR master mix.
3. Keep the reaction plates on ice until you are ready to load them into the 7300/7500
system.