Display Workstation
25
Show Information
Information related to the images can be shown/hidden as an overlay on
the images by selecting/deselecting it from the menu, as it is a toggle style
menu.
The displayed information is read from the DICOM header for each
image, and is different for each modality, such as CT, MRI, CT, etc. For
example, for MR images, the TI, TR, and TE values are displayed; while
for CT images, the KV and mAs values are displayed instead of TI, TR,
and TE.
ƒ Patient Orientation
The following image shows a description of the basic information that is
displayed for an image from the DICOM header.
This indicates the position and
orientation of the image in
relation to the patient when the
image was captured.
Modality Name
Exam Date
Exam Time
SP: Slice Location
SL: Slice Thickness
GT:Gantry/Detector Tilt
ST: Expose Time
TI: Inversion Time
TR: Repetition Time
TE: Echo Time
mAs: X-ray Tube Current
KV: KVP
Series Number
Image Number
Type of Compression
Place Name
Patient Name
Patient ID
Patient Sex
Acquisition Number
Scale Bar
Patient Orientation
Patient Orientation
Z: Zoom
W: Window Width
C: Window Level
Description of image related information
To show, or hide, image information in the Image Display Window,
8 Press [F9], or
8 Choose Display > Show Information from the menu, or
8 Click the Show Information button in the Toolbox.